Core Technologies

Alarm and Safety Function

Alarm

Alarm and Safty Function
Event
Visual
Audible
Signal to Alarm Contact
Status Alert Abnormal ambient (too high or too low), or abnormal freezer loading (too much warm product at once) STATUS lamp flashes None None
High Temperature Interior chamber warms beyond high-temperature setpoint ALARM lamp flashes; digital display flashes actual chamber temperature Intermittent tone; time delay of 15 minutes after reaching alarm setpoint to avoid false alarms. Yes
Low Temperature Interior chamber cools below low-temperature setpoint ALARM lamp flashes; digital display flashes actual chamber temperature Intermittent tone; time delay of 15 minutes after reaching alarm setpoint to avoid false alarms. Yes
Power Failure Loss of power ALARM lamp flashes; digital display flashes actual chamber temperature Intermittent tone Yes
Control Temperature Sensor Failure Sensor probe disconnect, short-circuit or failure ALARM lamp flashes Solid tone No
Filter Sensor Failure Sensor probe disconnect, short-circuit or failure ALARM lamp flashes Solid tone No
Interstage Refridgeration Sensor Failure Sensor probe disconnect, short-circuit or failure ALARM lamp flashes Solid tone No
Filter Check Clogged or dusty condenser filter FILTER lamp flashes None No
AutoReturn Touchpad key is not pressed for 90 seconds Reverts to chamber temperature display None No
Door Alarm Door Open OPEN lamp on Solid tone sounds after 2 minues (adjustable) Yes

 

 